How to set up data transmission
To enable scheduled data transmission, an HTTP or FTP server has to be specified in at least one of four available server configurations (Data transmission 1, Data transmission 2, Data transmission 3 or Data transmission 4). By default, the Measurement Data Server (MDS) of SOMMER Messtechnik is configured. If you have subscribed to this service, you can access the transmitted data on the server’s webpage.
Follow the steps below to set up regular data transmission:
- Activate mobile communication as described in How to activate mobile communication.
-
Set the parameters in the Data transmission menus as in the following examples:
EXAMPLEData transmission to an FTP server
Parameter
Value
Comment
ftp
00:30:00
every 30 minutes
00:00:17
random offset to avoid transmission conflicts to the server
data
optional: camera to transfer images
ftp.myserver.at
myuser
********
/data
an optional sub-directory
21
active
optional: passive
EXAMPLEData transmission to an HTTP server
Parameter
Value
Comment
http
00:10:00
every 10 minutes
00:00:31
random offset to avoid transmission conflicts
data
optional: camera to transfer images
mds.sommer.at
/Web-Service/sommerDaten.php
the data destination
80
NOTE At each interval, data since the last successful transmission are sent to the server. A copy of the data remains on the data logger until overwritten by newer data.
NOTE If measurement data and camera images need to be transmitted, two separate data transmission tasks need to be configured.