Flow velocity SV

The SV radar sensor is a measurement device for the contact-free determination of the flow velocity of open rivers and channels. The sensor detects the surface flow velocity by the principle of the Doppler frequency shift.

Sensor database ID

115

Name

SV

Measured Quantities

Flow velocity

Manufacturer

Sommer Messtechnik

Measurement range

 

Unit

Min

Max

Flow velocity

m/s

0.1

15

Analog setup

Wiring

ANALOG settings

Signal class

Current

Analog inputs

 

AIN signal type

4...20 mA

Warm-up time

0.1 sec

ADC - conv. Rate

62 Hz

ADC filter

Off

Sensor supply

 

SW2

Off

12 V / 14 V

Always on

5 V

Off

2.5 V

Off

External sensor supply

Not used

Additional measurement settings

 

12 V / 14 V output voltage

12 Volt

Measurement table

Function*

Identifier

Unit

Decimals

Scale

Offset

S-TYP

S-NUM

S-MEA

S-ADD

Actual

Velocity

m/s

2

0.3125**

-1.25**

AIN

 

AN 1

 

*options: off, actual, meanval, minimum, maximum, sum, intens., diff.; see MRL manual for details

**for a measurement range of 0 ... 5 m/s (can be set in SV setup)

 

SBP (RS-485) setup

 

Measurement table

Function*

Identifier

Unit

Decimals

Scale

Offset

S-TYP

S-NUM

S-MEA

S-ADD

Actual

Self-check

-

As S

 

0

SBP

1

1

 

Actual

Velocity

m/s

As S

 

0

SBP

1.

3

 

Actual

Quality (SNR)

-

As S

 

0

SBP

1

4

 

 

...**

 

 

 

 

 

 

 

 

*options: off, actual, meanval, minimum, maximum, sum, intens., diff.; see MRL manual for details

**Additional variables may be added according to the SV manual

SDI-12 setup

 

Measurement table

Function*

Identifier

Unit

Decimals

Scale

Offset

S-TYP

S-NUM

S-MEA

S-ADD

Actual

Self-check

-

As S

 

0

SDI12

0

1

 

Actual

Velocity

m/s

As S

 

0

SDI12

0

3

 

Actual

Quality (SNR)

-

As S

 

0

SDI12

0

4

 

 

...**

 

 

 

 

 

 

 

 

*options: off, actual, meanval, minimum, maximum, sum, intens., diff.; see MRL manual for details

**Additional variables may be added according to the SV manual