Analog measurements
With the MRL‑8E single ended voltage signals in the range of 0 to 2.5 V can be measured. The analog inputs AN3 and AN4 can also be configured to measure resistive sensors, e.g. Pt1000.
|
Option |
Description |
Terminals |
|---|---|---|
|
Voltage 2.5 |
Single ended voltage input 0 … 2.5 V |
AN1 ... AN4 |
|
NTC |
Measures the temperature of a NTC thermistor |
AN3 |
|
R meas > 2k |
Measures a resistance > 2 kΩ |
AN3 |
|
Pt1000 |
Measures the temperature of a Pt1000 temperature sensor |
AN4 |
|
R meas > 1k2 |
Measures a resistance > 1.2 kΩ |
AN4 |
|
R meas < 1k2 |
Measures a resistance > 0 Ω (preferably < 1.2 kΩ) |
AN4 |
|
Voltage 0.3 |
Single ended voltage input 0 … 0.3 V |
AN4 |
|
Voltage 2.5 |
Differential voltage input 0 … 2.5 V |
AN4 |
The signals received from a sensor are wired in the MRL‑8E as illustrated in Figure 8.
Figure 8 Internal wiring of analog MRL‑8E inputs
Principals
The illustration below shows a sequence of two measurement cycles of three analog inputs (AN1, AN2 and AN3) and switched power supply.
Figure 9 Principal of analog measurements
At the start of each measurement interval the sensor power supply is switched on. After the specified Warm-up time each of the three analog inputs is measured sequentially at the sampling rate defined in ADC - conv. Rate (red lines). After each measurement has been completed, the sensor power supply is switched off.
If ADC filter is active, the input is measured 3 times and the data logger returns the statistic specified in ADC filter.
The measurement interval of all sensors is specified in Measurement Interval.
